3200iSeries Indicators
Part No HA029006 Issue 5.0 Sept-12 63
13.5 Calibration Parameters
The following table lists the parameters available in the Calibration List.
CALIBRATION PARAMETER LIST cA L
Name Scrolling
Display
Parameter
Description
Value Default Access
Level
UCAL USER
CALIBRATION
To calibrate the 5
point linerisation
table.
Not 32h8i/SG
IDLE
L3 only
PnT.1
PnT.2
PnT.3
PnT.4
PnT.5
rset
PHASE CALIBRATION
PHASE
To calibrate low and
high offset
Fact Return to factory settings Fact
Factory
3.V.Hi High volts output from output 3
3.V.Lo Low volts output from output 3
3ma.H High mA output from output 3
3ma.L Low mA output from output 3
none
Not selected
none
Conf only
These
parameters
do not
apply to
32h8i/SG
0 Select mV i/p low calibration point
50 Select mV i/p high calibration point
0V Select V i/p low calibration point
10V Select V i/p high calibration point
150r Select PRT i/p low cal point
400r Select PRT i/p high cal point
CJC Select CJC calibration
GO CALIBRATION
START
To start the
calibration
sequence
NO Initial state NO
Conf
Yes Start
Busy Calibrating
Pass Calibration successful
faiL Calibration unsuccessful
SG.TYP STRAIN
GAUGE
CALIBRATION
TYPE
Selects the
calibration for the
sensor in use
SHnt
4-wire bridge type pressure
transducer
CeLL
L2
These three
parameters
are only
available in
32h8i/SG.
They are
used to
calibrate to
the strain
gauge
sensor.
See section
5.3.
CELL
Load cell
Comp
Comparison
SHUNT SHUNT
CALIBRATION
To set the high
calibration point for
the pressure
transducer in use
Off or 40.0 to 100.0
Off
LO.CAL STRAIN
GAUGE LOW
CAL
Calibrate the low
point
NO
Initial state NO
Yes
Start
HI.CAL STRAIN
GAUGE HIGH
CAL
Calibrate the high
point
Busy Calibrating
Pass Calibration successful
faiL Calibration unsuccessful
AUT.SG STRAIN
GAUGE
AUTO CAL
When selected this
perform an
automatic
calibration to the
strain gauge sensor.
See section 5.3.5.
NO Initial state NO
YES Start auto calibration
Lo These parameters automatically
appear as the calibration takes
place.
Hi
pass
faiL